David L. Constans
Gossman
Consulting, Inc.
ABSTRACT
This paper discusses the requirements of the HWC MACT CMS Performance Evaluation Plan and provides a model plan. This evaluation plan must be submitted with the ICPT plan and is to specify the inspection, maintenance and external quality control of the monitoring system components; thermocouples, D-P cells, opacity monitors and CEMs. Comments from the EPA on submitted plans will be included and discussed if these comments are available prior to the conference paper submittal date.
The HWC MACT rule
requires the submittal of a Continuous Monitoring Systems (CMS)
Performance Evaluation Test Plan with the submittal of the Initial
Comprehensive Performance Test Plan (40CFR 63.1207(e)).
The CMS Performance Evaluation Test Plan requirements are
detailed in 40CFR 63.8(e). Also 40CFR
63.1207(d) requires that the quality control program outlined in 4CFR
63.8(d) also be included in the performance evaluation.
To complicate things somewhat the HWC MACT regulation includes
in 40CFR 63.1209(a), (b) and (c) requirements for specific parts of the
CMS system. To complicate things still
further the EPA issued a technical amendment to the HWC MACT (Nov. 9,
2000 volume 65, number 218, page 67268-67272). Part
of the amendment was to clarify the difference between a CMS Evaluation
Plan and a CMS Evaluation Test Plan and the requirement to submit only
the CMS Evaluation Test Plan with the Initial Comprehensive Performance
Test Plan.
In the amendment the
agency has defined the CMS Evaluation Test Plan to include “testing
necessary to demonstrate calibration, minimization of malfunction, and
how the CMS will meet the performance specification.”
The CMS Evaluation Plan is to include the quality assurance
programs that “specified how a source would maintain calibration...and
minimize malfunction.” It would appear
that the CMS Evaluation Test Plan is to exclude QA programs, yet
somehow “minimize malfunctions” and “meet...performance specifications.” Both of which would require such programs. Also 40CFR 63.8(e) defines CMS Evaluation Test
Plan as including the QA programs. The
only resolution that readily appears is to include the QA programs but
without specifying the time intervals of the activities.
It had been hoped that one or more of the three CMS Evaluation Test Plans written by GCI for our clients and submitted to the regional EPA, would have been reviewed by that authority, and such review included as an additional guide in preparing this paper. Unfortunately this has not occurred. Consequently this paper presents a format for a CMS Evaluation Test Plan that has not been critiqued by the agency.
The CMS Evaluation
Test Plan has been reduced to two sections.
A master table that provides a listing of the monitored parameters, the
device used for such monitoring, the range of measurement, accuracy of
measurements (data quality objectives), internal QA and external QA
activities, record locations, and comments. See
attachments.\
The second section
includes subsections for each monitored parameter.
A subsection includes a table repeating the first four columns
of the master table for each of the monitored parameters and text
describing the internal and external QA activities and any routine
maintenance performed on the device.
Parameter
Monitored:
These are the parameters that must be continuously monitored as specified in the HWC MACT permit.
Device
Description:
This is the manufacturer and model number of the measurement device.
Range and Units
of Measurement:
This should be the range the recorder attached to the device and the units of measurement of that recorder. The actual range of the measurement device may be larger but exceeding the range of the recorder is a violation of the regulation.
Accuracy of
Measurement (DQOs):
This is a value expressing the allowable inaccuracy of the measurement device. In some instances, these values are dictated by the Performance Specification (PS1 (Opacity Monitor), PS4B (Oxygen Monitor) or PS8A (THC Monitor)) in other instances it is based on the inherent capabilities of the instrument and/or the sampling/analysis procedure. In any event this is the accuracy which the device’s recorder must achieve upon calibration or during an audit.
Table 1. Master Table
Continuous monitoring systems performance evaluation summary
for a typical preheater/precalciner cement production facility
Monitored Parameter |
Device Description |
Range and Units of Measurement |
Accuracy of Measurement (DQOs) |
Internal Quality Assurance
Activities |
External Quality Assurance
Activities |
Location of Records |
Comments |
Kiln Feed
Weigh Feeder |
ABC Company
Weigh Feeder |
0-250 tons
per hour |
+/- X% of actual
weight |
Continuous
monitoring of function. Inspection as
required. |
Calibration |
Alternate
Fuels Office |
|
Pumpable
Hazardous Waste Feed |
XYZ Mfg.
Mass Flow Meter |
0-250 pounds
per minute |
+/- X% of actual
weight |
Continuous
monitoring of function. Inspection as
required. |
Calibration |
Alternate
Fuels Office |
|
Calciner
Coal Feed Weigh Feeder |
ABC Company
Weigh Feeder |
0 - 20 tons
per hour |
+/- X% of actual
weight |
Continuous
monitoring of function. Inspection as
required. |
Calibration |
Alternate
Fuels Office |
|
Kiln Coal
Feed Weigh Feeder |
DEF Co.
Rotor Weigh Feeder |
0 - 15 tons
per hour |
+/- X% of actual
weight |
Continuous
monitoring of function. Inspection as
required. |
Calibration |
Alternate
Fuels Office |
|
Tire Chip
Weigh Feeder |
GHI Inc.
Belt Scale |
0-10 tons
per hour |
+/- X% of actual
weight |
Continuous
monitoring of function. Inspection as
required |
Calibration |
Alternate
Fuels Office |
|
4th
Stage Temperature |
Generic Type
K thermocouple |
1400-2200
°F |
+/- X% of range |
Continuous
monitoring of function. Inspection as
required |
Calibration |
Alternate
Fuels Office |
|
Main
Baghouse Inlet Temperature |
Generic Type
K thermocouple |
0-1100
°F |
+/- X% of range |
Continuous
monitoring of function. Inspection as
required |
Calibration |
Alternate
Fuels Office |
|
By-Pass
Baghouse Inlet Temperature |
Generic Type
K thermocouple |
0-800
°F |
+/- X% of range |
Continuous
monitoring of function. Inspection as
required |
Calibration |
Alternate
Fuels Office |
|
Kiln Hood
Pressure |
Generic DP
Cell Pressure Transmitter |
-1.0 to
+0.25 inches of water column |
+/- X% of range |
Continuous
monitoring of function. Inspection as
required |
Calibration |
Alternate
Fuels Office |
|
Kiln Rear
Pressure |
Generic DP
Cell Pressure Transmitter |
0 to –3
inches of water column |
+/- X% of range |
Continuous
monitoring of function. Inspection as
required |
Calibration |
Alternate
Fuels Office |
|
Main
Baghouse Differential Pressure |
Generic DP
Cell Pressure Transmitter |
0 to 20
inches of water column |
+/- X% of range |
Continuous
monitoring of function. Inspection as
required |
Calibration |
Alternate
Fuels Office |
|
By-Pass
Baghouse Differential Pressure |
Generic DP
Cell Pressure Transmitter |
0 to 20
inches of water column |
+/- X% of range |
Continuous
monitoring of function. Inspection as
required |
Calibration |
Alternate
Fuels Office |
|
Main Stack
Opacity Monitor |
JKL
Instruments Model XXX |
0 – 100% |
Zero/Upscale
drift (24hr) - £ 2% Opacity Calibration
Error - £ 3% Opacity Pass 7 day
operational test period Response time £ 10 seconds Data
recorder resolution £ 0.5% |
Zero
calibration Upscale
calibration, and for
each, a 24 hour drift determination. |
A
Performance Evaluation consisting of the following tests: a calibration
error, a response time determination, 7
day operational test period with a daily
24 hour zero/upscale calibration drift test. |
Daily
Calibration Drift records on monitor data terminal. Annual
Performance Evaluation records in the Alternate Fuels Office |
|
Total
Hydrocarbon Monitor (on By-Pass Duct) |
MNO
Environmental Instrument Company Model ZZ |
0-100 ppmv
as propane |
Zero/Upscale
drift (24hr) - ± 3 ppm Calibration
Error - ± 5 ppm Pass 7 day
operational test period Response time £ 2 minutes Data
recorder resolution £ 0.5ppm Permit
requires accuracy of: ±0.5% of calibrated Range |
Zero
calibration Upscale
calibration, and for
each, a 24 hour drift determination. |
A
Performance Evaluation consisting of the following tests: a calibration
error, a response time determination, 7
day operational test period with a daily
24 hour zero/upscale calibration drift test. |
All
documentation is kept in the CEM room operational record |
Two monitors
are used for this service. Both are
calibrated each day. |
Oxygen
Monitor (on the By-Pass Duct) |
PQR Ind. Model YYY |
0-25% |
Zero/Upscale
drift (24hr) - £ 0.5% O2 Calibration
Error - £ 0.5% O2
Pass 7 day
operational test period Response time £ 2 minutes Data
recorder resolution £ 0.5% O2 Permit
requires accuracy of: ±0.5% of calibrated Range |
Zero
calibration Upscale
calibration, and for
each, a 24 hour drift determination. |
A
Performance Evaluation consisting of the following tests: a calibration
error, a response time determination, 7
day operational test period with a daily
24 hour zero/upscale calibration drift test. |
All
documentation is kept in the CEM room operational record |
Two monitors
are used for this service. Both are
calibrated each day. |
Trace Metal Input Control |
Sample
collection methods and sample analysis methods are covered in the Waste
Analysis Plan (WAP). The WAP and the
laboratory SOPs detail the QA and QC requirements.
|
Internal
Quality Assurance Activities:
These activities are stated in 40CFR63.8(e)(3)(ii) to include, activities planned by routine operators and analysts to provide an assessment of CMS performance. All of the parameter monitoring devices send a signal to the central processing unit where the values are monitored for compliance with the permit requirements. In effect the computer monitors the function of the devices and can, and often is programmed to sound an alarm if the device fails to function, that is fails to send a signal within certain parameters. Additionally, experienced operators would quickly notice device output readings that were inconsistent with values of other operational parameters. An alarm or noted inconsistency would lead to inspection of the device and if required maintenance. The table provides a summary of these activities. The subsections in the QA section provide more detail.
External Quality Assurance Activities:
The regulation in 40CFR63.8(e)(3)(ii) states that the external QA program include: “system audits that include the opportunity for onsite evaluation by the Administrator of instrument calibration, data validation, sample logging, and documentation of quality control data and field maintenance activities”. These activities are generally performed at specified intervals, either dictated by a performance specification (such as PS1 for Opacity Monitors, PS4B for the Oxygen monitor or PS8A for the THC monitor) by the device’s manufacturer or by good engineering practice based on site specific conditions. For the purposes of the submittal of a CMS Evaluation Test Plan these frequencies are not specified. The table provides a summary of these activities. The subsections in the QA section provide more details.
Location of
Records:
This is self-obvious. These records may be paper or electronic, the regulation is not specific about this. Certain records such as a Relative Accuracy Performance Test executed by a contraction would of course be paper files. Internally generated, such as the one-minute values that are subsequently used to calculate hourly rolling averages will of course be in electronic format.
Comments:
If
there are
redundant or standby monitors or devices this should be noted here.
The monitored parameter QA programs section of the
test plan provides additional details on internal, and external QA
activities and any routine maintenance activities.
See the Example QA Programs below. These
QA programs will frequently include attachments.
Monitored
Parameter |
Device Description |
Range
and Units of
Measurement |
Accuracy
of Measurement (DQOs) |
Pumpable Hazardous Waste Feed |
ABC Company Mass Flow Meter |
0-250 pounds per minute |
+/- X% of actual weight |
No routine maintenance required.
Manufacturer’s literature is attached on the mass meter as well as a blank mass flow rate device accuracy deterermination form. The records of this test and, if necessary, subsequent calibrations are kept in the Alternate Fuels office.
Monitored
Parameter |
Device
Description |
Range and Units of
Measurement |
Accuracy
of Measurement (DQOs) |
Main Stack Opacity Monitor |
JKL Instruments Model XXX |
0 – 100% |
Zero/Upscale drift (24hr) - £ 2% Opacity Calibration Error - £ 3% Opacity Pass 7 day operational test period Response time £ 10 seconds Data recorder resolution £ 0.5% |
The device has an automatic calibration
feature provided by the manufacturer. This
allows the device to conduct its own daily zero and upscale calibration
and to perform a 24 hour drift calculation. Failure
to perform the calibration or a drift in excess of 2% opacity would
result in an alarm and the operator would initiate an investigation by
maintenance. Diagnosis and repair would
follow. Since the regulation allows ten
seconds between samples, this allows the device to execute hourly
“dirty window” and calibration corrections as well.
A “Dirty Window” exceedance warning will be issued by the device
if a calibration fails this would be followed by inspection and
maintenance as required.
The opacity windows are inspected daily and are cleaned as needed.
Annually, the device is subjected to inspection and calibration per the requirement of PS-1 as detailed in the Performance Specification Verification Procedure Section 7. A copy of PS-1 is attached, including blank forms and calculations. Briefly, the procedure consists of the following:
a) There is a seven-day conditioning period to determine that the device is operating well enough to be tested. For retesting this is unnecessary.
b) The device is subjected to a calibration error test over three points in its range: low, midrange, and upscale. The device must exhibit a £3% opacity difference as calculated in the method.
c) The device is subjected to the response time test and must exhibit a response time of £10 seconds.
d) The device must then pass the seven day operational period where no repairs are allowed to the device and the device must successfully calibrate daily and exhibit £2% opacity drift each 24 hours.
e) The Data Recorder Resolution is
documented to be £0.5%
opacity.
The daily calibration and calibration drift calculation is recorded in the Data Log in the Opacity Device Terminal. The Annual Performance Specification Verification documentation is kept in the Alternate Fuels office. Maintenance repair records are kept in the Electrical Department, such as: manufacturer’s literature on calibration procedures, installation procedures and maintenance procedures or copies of the applicable performance specification. These attachments may also have site specific procedures written by facility management, such as special calibration procedures or maintenance procedures, this may include special forms for calculating instrument accuracy or for reporting maintenance on the device.
The device subsection in the QA programs section of the test plan provides additional details on internal, and external QA activities and any routine maintenance activities. These subsections will frequently include attachments such as: manufacturer’s literature on calibration procedures, installation procedures and maintenance procedures or copies of the applicable performance specification. These attachments may also have site specific procedures written by facility management, such as special calibration procedures or maintenance procedures, this may include special forms for calculating instrument accuracy or for reporting maintenance on the device.
The regulation in
40CFR63.2(definitions) defines CMS to include “manual or automatic
monitoring that is used for demonstrating compliance with an applicable
regulation on a continuous basis.”
Since the regulation
requires that trace metal feed rate limits not be exceeded, it is
necessary to include trace metals analysis as proof of the CMS
Evaluation Test Plan. The trace metals
analysis of the kiln feeds are entered into a data base which is used
to compute trace metals feed rates based on the mass input of the
various feeds. The method and QA/QC
requirements for these trace metals analyses are already documented in
the facilities waste analysis plan (WAP) and it would be redundant to
include the WAP with the CMS Evaluation Test Plan, it should however be
referenced in the Test Plan.
This format for a CMS Evaluation Test Plan should adequately address the requirements. The extent of the additional details deemed necessary by the agency is currently unknown. However, this format readily allows insertion of necessary documentation. As noted earlier the difference between the Evaluation Test Plan and the Evaluation Plan is the lack of QA activity frequency. This can easily be added to the Internal QA and External QA columns in the master table.